CF InPlex ASR Card; Catalog Number: 95-0501.

Company
Hologic, Inc
Recall Initiated
March 31, 2016
Terminated
August 9, 2016
Recall Number
Z-1579-2016
Quantity
2580 total
Firm Location
San Diego, CA

Reason for Recall

Complaints of false positive HET mutations and leaking cards; could lead to unnecessary invasive testing during pregnancy and the associated risks of such invasive testing.

Distribution

Distributed in the states of OH, AL, IL, AZ, AR, NJ, TX, GA, CO, DE, NC, MI, FL, ME, CA, IN, OR, KS, KY, MO, WA, NY, OK, WV, CT, IA, LA, RI, MA, and PA, and in the country of Canada.

Lot / Code Info

Lot Numbers: U34xx U35xx (xx denotes sublots)

Root Cause

Nonconforming Material/Component

Action Taken

Customer notification letters were sent 03/31/16. Customers were instructed to immediately discontinue the use of the affected lot numbers. If a HET or MUT result was reported with the affected lots, Hologic recommends that customers notify the recipient of the result and request the patient be retested with another test method. Any HET or MUT results that were previously retested and reported out as negative do not require further testing. The firm requests that customers segregate any remaining inventory of all affected lots listed on the Customer Response Form. After segregation is complete reconcile the number of units (ASR Card or IVD Molecular Test Kit) which remain in inventory upon receipt of notification against the number originally shipped to site, and destroy all remaining inventory of the affected lots. Document the following in the appropriate fields on the Customer Response Form: 1. If have zero of the affected products on hand check the first box. 2. If have some inventory of the affected products on hand check the second box and then document the quantity on hand below, in the Qty On-Hand column. Note: When counting the number of units remaining in inventory count each unit (ASR Card box or IVD Molecular Test Kit), whether partial or full, as a unit. 3. Print name, phone and email address, in case the firm needs to follow up. 4. Sign and date the form. Signing acknowledges compliance with this recall notification, including destroying all remaining inventory of the affected products. When this documentation is complete fax the CRF to Hologic Technical Support using fax number: 858-410-8250. Alternatively, you can scan the document(s) and email to Hologic Technical Support at molecularsupport@hologic.com. If laboratory has shared any of the affected cards noted in the CRF with another laboratory contact Hologic Technical Support immediately using the phone number below, so that Hologic may contact the other laboratory t

More recalls by Hologic, Inc

View all recalls by this company →