Infinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplate reader with injector option. The Infinite F500 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite F500 is intended as a general purpose laboratory instrument for professional use, supporting common microplate conforming to the ANS/SBS standards.

This recall has been terminated (originally issued July 2, 2015).

Company
Tecan Us, Inc.
Recall Initiated
May 29, 2015
Posted
July 2, 2015
Terminated
August 18, 2017
Recall Number
Z-1967-2015
Quantity
868 in total
Firm Location
Morrisville, NC
Official Source
View on FDA website ↗

Reason for Recall

A firmware bug causes incorrect values to be displayed for specific measured luminescence values.

Distribution

US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC

Lot / Code Info

Material Number 30019337.  Serial numbers 1212004416 1009002605 612000003 811002510 712004158 906000965 902005109 912003472 1103002583 1411007090 704000003 702000001 708000500 1003004146 1109000103 1412005124 805001219 911005026 708005532 705007195 810000248 1010001663 712004152 708005533 1412005709 1103003780 706004455 608000002 908000185 1009002604 1308001035 804005914 1311007600 804000984 705009034 1309003195 801008055 706007961 901006341

Root Cause

Device Design

Action Taken

Tecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk.