MC-500 Multicolor Laser Photocoagulator; Indicated for use in retinal photocoagulation for treatment of ocular fundus diseases like diabetic retinopathy, age-related macular degeneration, retinopathy of prematurity and retinal detachment.

This recall has been terminated (originally issued February 19, 2015).

Company
Nidek Inc
Recall Initiated
March 27, 2012
Posted
February 19, 2015
Terminated
February 19, 2015
Recall Number
Z-1127-2015
Quantity
16 units in commerce
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

The spot size control of the scan delivery unit of the MC-500 Multicolor Laser Photocoagulator may turn by itself.

Distribution

Nationwide Distribution- including the states of CA, MI, NY, IL, NH, KS, NY, UT, IA, WI, TX, and FL.

Lot / Code Info

Serial numbers: 20023,20039, 20040, 20042, 20047, 20049, 20050, 20051, 20052, 20062, 20063, 20064, 20066, 20067, 20068, 20076

Root Cause

Component design/selection

Action Taken

Nidek Japan issues a Engineering Change Order and Nidek reps visited affected sites to perform correction.