Nidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII. Opthalmic Device. The system is composed as a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for system control.

This recall has been terminated (originally issued November 17, 2014).

Company
Nidek Inc
Recall Initiated
August 6, 2012
Posted
November 17, 2014
Terminated
November 17, 2014
Recall Number
Z-0223-2015
Quantity
14 units
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

Contact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event for the patient.

Distribution

US Distribution including the states of CO, NY, MO, LA, ME, CA, MI, AZ, CT and FL.

Lot / Code Info

Serial numbers: 530054 530055 530070 530073 530078 530082 530086 530087 530098 530106 530121 530123 530150 530155

Root Cause

Component design/selection

Action Taken

Nidek Engineers visited each site to perform correction - fuse holders have been inspected and replaced if necessary. An Engineering Change Order was issued on June 1, 2012, for inspection and/or replacement of affected fuse holders by Nidek Service Engineers. For further questions please call (510) 353-7785.