RT-5100 Refractor and RT-3100 Refractor; Distributed by MARCO Opthalmic: Manufactured by NIDEK CO., Ltd., Aichi, Japan. Instrument provides means of positioning spherical and cylindrical lenses, prisms and other optical devices in front of a subject's eyes for purpose of determining refractive error and binocular functions.

This recall has been terminated (originally issued June 23, 2014).

Company
Nidek Inc
Recall Initiated
June 9, 2014
Posted
June 23, 2014
Terminated
October 30, 2015
Recall Number
Z-1853-2014
Quantity
~4,285 units RT-5100; 167 units of RT-3100
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

Near Point Chart Arm of RT 5100 and RT-3100 Refractors may lower spontaneously with potential for patient injury.

Distribution

US distribution only.

Lot / Code Info

RT-5100 and RT-3100, various serial numbers.

Root Cause

Use error

Action Taken

Recall/Field Correction letters were sent to doctors on June 9, 2014 with a preventive maintenance instruction to be performed on the device by user. Also included are the necessary tools to perform actions. Customers are to complete and return the Recall Response Card. Questions are to be directed to Customer Service. Contact Information: Customer Service: 1-800-891-1772 E-mail: RT-5100 update@noritsu.com