CARTO 3 System Interface Cable, 34-pin to 10-pin, Red, 10' - Part Number: D-1286-18; Catalog Number CR3410CT This product provides a signal interface between a Biosense Webster electrophysiology catheter and the appropriate equipment. This cable may be re-used subject to established cleaning and sterilization restrictions
- Company
- Biosense Webster, Inc.
- Recall Initiated
- October 5, 2010
- Posted
- January 7, 2011
- Terminated
- May 7, 2012
- Recall Number
- Z-0884-2011
- Quantity
- 1692
- Firm Location
- Irwindale, CA
Reason for Recall
When connecting this cable to the mapping system, loss of Body Surface ECG signals may occur. This may result in inaccurate locations appearing for catheters displayed 3D on the CARTO monitor.
Distribution
Worldwide Distribution: USA, including the states of: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, and WV, and the countries of Belgium, China, Israel, Japan, Singapore, and Thailand.
Lot / Code Info
Part Number: D-1286-18; Catalog Number CR3410CT
Root Cause
Device Design
Action Taken
Biosense Webster issued an Urgent Field Safety Notice letter dated October 5, 2010. The letter identified the product, the problem,and the action to be taken by the customer. Customers were instructed to return the affected cable, accompanied by the attached Return Authorization form. For questions regarding this recall call Biosense Webster at 909 839-8874.