RecallDepth
Class II Terminated

Nutropin AQ NuSpin 20 (somatropin (rDNA origin) injection} 20 mg/ 2mL Genentech, South San Francisco, CA 94080, NDC#50242-0076-01

Company
Genentech, Inc.
Recall Initiated
September 20, 2012
Posted
December 12, 2012
Recall Number
D-087-2013
Voluntary / Mandated
Voluntary: Firm initiated
Quantity
10,306 syringes
Firm Location
South San Francisco, CA, United States

Reason for Recall

Miscalibrated and/or Defective Delivery System: Genentech has received complaints for Nutropin AQ NuSpin 10 & 20 reporting that the dose knob spun slowly and the injection took longer than usual (slow dose delivery).

Distribution

Nationwide

Lot / Code Info

Lot# 936674 Exp. 09/30/13

More recalls by Genentech, Inc.

View all recalls by this company →