RecallDepth
Class II Terminated

LifeGas OXYGEN COMPRESSED UN1072 USP, Distributed by: Linde

Company
Linde Gas Llc
Recall Initiated
May 4, 2012
Posted
June 27, 2012
Recall Number
D-1383-2012
Voluntary / Mandated
Voluntary: Firm initiated
Quantity
283 cylinders
Firm Location
Columbus, OH, United States

Reason for Recall

CGMP Deviations: The pressure gages, vacuum gages, and thermometer had surpassed the calibration expiry period, which may have resulted in overfill/underfill of oxygen cylinders.

Distribution

Distributed only within Ohio. No Federal Government, foreign or Canadian consignees received product.

Lot / Code Info

LOT#: 117H201 LOT#: 117H202