ConforMIS iUni ( Unicondylar Knee System ) Left Medial Implant Kit Product Code: M57220600010. For cemented use only. System consists of femoral and tibial components. The femoral component is CoCrMo personalized on its sub-chondral and articular surfaces to match patient's anatomy. For use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface in patients with evidence of inadequate healthy bone for support of the implanted component.

This recall has been terminated (originally issued July 15, 2009).

Company
Conformis Inc
Recall Initiated
April 23, 2009
Posted
July 15, 2009
Terminated
October 14, 2009
Recall Number
Z-1622-2009
Quantity
10 Units
Firm Location
Burlington, MA
Official Source
View on FDA website ↗

Reason for Recall

Microscopic cracks potential for pre-mature revisions in patients implanted with the system.

Distribution

Worldwide Distribution - US including states of CA, LA, MI and TX and Germany.

Lot / Code Info

Serial Numbers: 2565, 2686, 2685, 2308, 2739, 2728, 2299, 2726, 2681 and 2521.

Root Cause

Process control

Action Taken

ConforMIS contacted accounts and presented the surgeon with the customer communication letter dated April 23, 2009, customized for the surgeon by serial number and patient identification. Product not yet implanted will be recovered. A patient evaluation and monitoring protocol will be recommended for potentially affected product already implanted. Contact ConforMIS for further questions at 1-781-345-9164.