Biomet Biolox-Delta Modular Ceramic Head, 28 mm head diameter, plus 5 neck, type 1, taper, sterile, Biomet Orthopedics, Inc., Warsaw, IN; Catalog Number 12-115112. A hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint.

This recall has been terminated (originally issued July 14, 2009).

Company
Biomet, Inc.
Recall Initiated
March 16, 2009
Posted
July 14, 2009
Terminated
October 6, 2010
Recall Number
Z-1613-2009
Quantity
174
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The firm has received reports of fractured heads, requiring revision surgery.

Distribution

Worldwide Distribution -- US including PR and states of AZ, AL, AR, CA, CO, FL, IL, IN, KY, KS, MA, MI, LA, NY, SD, GA, MN, PA, NJ, MD, MI, OH, ME, TX, WA, WI and VA and countries of Australia, Belgium, Canada, Costa Rica, Finland, Netherlands and South Wales.

Lot / Code Info

Lots 114380, 175490, 229000, 310710, 342670, 385840, 403940, 467870, 484970, 572770, 641430, 705380, 761420 and 792280.

Root Cause

Process control

Action Taken

Biomet Orthopedics issued an "Urgent Medical Device Recall Notice" dated March 16, 2009 to users describing the affected device. Consignees were instructed to locate and discontinue use of the product and quarantine until a sales representative can arrange for removal. Consignees were also asked to complete and return the attached "Fax-Back Response Form" to Biomet at 1-574-372-1683. Further questions can be directed to Biomet Orthopedics at 1-800-348-9500.