RecallDepth

Immulite 2000 and Immulite 2500 digital Fluidics PC Board, Part Number 450116-0002. Performs chemiluminescent immunoassays using serum, plasma, and/or urine for in vitro diagnostic testing.

This recall has been terminated (originally issued May 26, 2009).

Company
Siemens Medical Solutions Diagnostics
Recall Initiated
January 5, 2009
Posted
May 26, 2009
Terminated
April 19, 2010
Recall Number
Z-1399-2009
Quantity
295 Digital Fludics PCB (88 IMMULITE 2000 and 2500 and 207 as spare parts).
Firm Location
Flanders, NJ
Official Source
View on FDA website ↗

Reason for Recall

Design Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.

Distribution

Worldwide Distribution to states including MA, NY, PA, MD, GA, FL, OH, IN, MI, IL, MO, LA, UT, AZ and CA and countries of Canada, Czech Republic, Germany, Poland, Hungary, Belgium, Russia, Republic of South Africa, Greece, Israel, Turkey, Spain, Portugal, Denmark, France, Great Britain, India, China, South Korea and Mexico.

Lot / Code Info

IMMULITE 2000: Serial Numbers M5080 to M5137; and IMMULITE 2500: Serial Numbers F0778 to F0818.

Root Cause

Process design

Action Taken

Siemens Healthcare Diagnostics Inc. issued an "Urgent: Device Recall" notice dated January 2009 mailed to all direct customers of the affected product. The letter described the problem, provided "work-around" instructions and also provided a Recall Response Form to be returned to the firm via fax. For further information, please contact your local Siemens Healthcare support representative or Siemens Healthcare Diagnostics Inc. directly at 1-973-927-2828.