RecallDepth

740 Series Multiparameter Monitor Software Version 3.0 Model Numbers: 740-1, 740M-1, 740-2T, 740M-2T, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS, 740-3NL, 740M-3NL, 740-3NN, and 740M-3NN CAS Medical Systems, Inc., Branford, Connecticut 06405 Indicated for use for non-invasive monitoring of blood pressure, oxygen saturation, pulse and temperature of adults, pediatric and neonatal patients, in the care of health professionals.

This recall has been terminated (originally issued September 16, 2008).

Company
Cas Medical Systems, Inc.
Recall Initiated
June 24, 2008
Posted
September 16, 2008
Terminated
August 5, 2010
Recall Number
Z-2203-2008
Quantity
820 units
Firm Location
Branford, CT
Official Source
View on FDA website ↗

Reason for Recall

Software: Audible alarm may be silenced if changed from the factory setting

Distribution

Worldwide Distribution --- including USA and countries of Canada, Jordan, UK, France, and Taipei.

Lot / Code Info

Serial Numbers: 0809834-0823035

Root Cause

Software change control

Action Taken

CASMED notified customers by a Product Correction Notification letter on 6/24/08 to advise of problem of the audible alarm and a software upgrade to correct the problem. The firm requests that the alarm volume not be adjusted from its factory set value to ensure that alarms sound correctly. Also, the letter states that if the volume needs to be adjusted, that customers needs to be sure that they only select two clicks above, or two clicks below the mid-point, and that there is a corresponding beep when you make your selection. Customers are asked to review the attached list of serial numbers that will need to be updated with new software and to then email the a contact at the firm once updates have been completed. All updates must be completed by August 31, 2008.

More recalls by Cas Medical Systems, Inc.

View all recalls by this company →