LS-202 Large LeadWear Combo Pack - Disposable ECG Monitoring component intended for use as a radiofrequency signal transmitter and receiver of electrocardiograph physiological signals which are displayed on the ECG monitors of various manufacturers' systems that have been validated for compatibility.

This recall has been terminated (originally issued September 16, 2008).

Company
Lifesync Corporation
Recall Initiated
December 13, 2007
Posted
September 16, 2008
Terminated
October 1, 2008
Recall Number
Z-1660-2008
Quantity
576 completed leadwire units
Firm Location
Ft Lauderdale, FL
Official Source
View on FDA website ↗

Reason for Recall

LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, intended for use as a radiofrequency signal transmitter and receiver of electrocardiograph physiological signals which are displayed on the ECG monitors of various manufacturers' systems that have been

Distribution

Class II Recall - Nationwide Distribution.

Lot / Code Info

All Lots

Root Cause

Nonconforming Material/Component

Action Taken

Firm issued a Device Recall Letter on 12/14/2007. Users were requested to remove the product from inventory and discontinue use of the Large Sized LeadWear with the 12 Lead Adaptors. The current stock of product in inventory will be held under a stop shipment, pending destruction of non-conforming product.