ARCHITECT Ferritin Reagent; in vitro diagnostic; list numbers 6C11-20 (4 x 100 tests), 6C11-25 (1 x 100 tests) and 6C11-30 (4 x 500 tests); Abbott Laboratories, Abbott Park, IL 60064 USA

Company
Abbott Laboratories Mpg
Recall Initiated
June 6, 2006
Posted
July 4, 2006
Terminated
January 18, 2008
Recall Number
Z-1204-06
Quantity
10,725 kits
Firm Location
Abbott Park, IL

Reason for Recall

Some ARCHITECT Ferritin Reagent lots may not be meeting the accuracy by correlation to AxSYM claims as they are listed in the package insert (slope of 1.0 +/- 0.2).

Distribution

Canada, Germany, Uruguay, Japan, Singapore, Hong Kong, South Korea, Australia and New Zealand

Lot / Code Info

list numbers 6C11-20, 6C11-25 and 6C11-30; all lots

Root Cause

Other

Action Taken

Abbott international affiliates were e-mailed copies of the recall letter on 6/6/06 for translation where needed and dissemination to their accounts receiving the affected product. The end users were advised that a shift in patient and/or control results occurs when changing to reagent and calibrator lots manufactured using the new internal reference standard, and that the accuracy by correlation between Architect and AxSYM Ferritin reagent lots does not match those stated in the package insert. The accounts advised to set new targets for their controls or use the enclosed targets in Attachment A when they change to the assay materials manufactured using the new internal reference standards. The accounts were requested to keep a copy of the letter on file until the revised package insert is received.