Medtronic Kinetra 7428. Dual Program Neurostimulator for Deep Brain Stimulation. Rx only. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. Kinetra neurostimulator is a component of the Activa System. The Medtronic Activa System is an implantable, multiprogrammable system that delivers electrical stimulation to selected areas of the brain. The dual program Model 7428 Kinetra Neurostimulator generates electrical signals that are transmitted to the brain. The neurostimulator consists of electronic circuitry and a battery, which are hermetically sealed in a titanium case.

This recall has been terminated (originally issued May 27, 2006).

Company
Medtronic Neurological
Recall Initiated
April 4, 2006
Posted
May 27, 2006
Terminated
December 26, 2006
Recall Number
Z-0861-06
Quantity
24
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

A specific subset of model 7428 Kinetra implantable neurostimulators may fail due to lifted wirebonds between the hybrid circuit and battery. This failure mechanism may present clinically as loss of functionality and a return of the patient's underlying medical conditions.

Distribution

USA--States of CA, OR, PA. Internationally -Belgium, France, Germany, Italy, Spain, Switzerland, United Kingdom

Lot / Code Info

NFD621879S, NFD621891S, NFD621901S, NFD621981S, NFD621976S, NFD622170S, NFD622185S, NFD622189S, NFD622201S, NFD622247S, NFD622279S, NFD623035S, NFD622345S, NFD622349S, NFD622372S, NFD622382S, NFD622387S, NFD622400S, NFD622414S, NFD622466S,  NFD622542S, NFD622564S, NFD622629S, NFD622664S

Root Cause

Other

Action Taken

An Urgent Device Recall Important Patient Management Information letter was sent to implanting and following physicians of patients who have received the device. The letter describes the issue and provides recommendations.

More recalls by Medtronic Neurological