Haemonetics Cell Saver 5 Autogolous Blood Recovery System Model Numbr: LN02005-110-E

This recall has been terminated (originally issued April 29, 2006).

Company
Haemonetics Corporation
Recall Initiated
December 21, 2005
Posted
April 29, 2006
Terminated
March 12, 2008
Recall Number
Z-0822-06
Quantity
3 units
Firm Location
Braintree, MA
Official Source
View on FDA website ↗

Reason for Recall

Device emits radiofrequency which exceed the IEC 60601-1-2 International Standard

Distribution

Nationwide

Lot / Code Info

Serial Numbers: 94H180CD-1; 95A048CS; 95J177CS

Root Cause

Other

Action Taken

Haemonetics notified accounts by Certified Letter on 12/21/05. A field correction will be schedule to replace the LCD .