Haemonetics Cell Saver 5 Autogolous Blood Recovery System Model Numbr: LN02005-110-E
This recall has been terminated (originally issued April 29, 2006).
- Company
- Haemonetics Corporation
- Recall Initiated
- December 21, 2005
- Posted
- April 29, 2006
- Terminated
- March 12, 2008
- Recall Number
- Z-0822-06
- Quantity
- 3 units
- Firm Location
- Braintree, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Device emits radiofrequency which exceed the IEC 60601-1-2 International Standard
Distribution
Nationwide
Lot / Code Info
Serial Numbers: 94H180CD-1; 95A048CS; 95J177CS
Root Cause
Other
Action Taken
Haemonetics notified accounts by Certified Letter on 12/21/05. A field correction will be schedule to replace the LCD .