XLie with Tubing, 12 cm, Electrosurgical Device-RITA * Medical Systems, Inc. * StarBurst Xli enhanced with tubing set * * REF/Catalog: 700-103027 * 4-7 cm Diameter * 12cm Length * Sterile/R * RITA * Manufactured by: Rita Medical Systems, Inc., One Horizon Way, Manchester, Georgia 31816, USA

This recall has been terminated (originally issued May 31, 2006).

Company
Rita Medical Systems, Inc.
Recall Initiated
May 2, 2006
Posted
May 31, 2006
Terminated
February 5, 2008
Recall Number
Z-0888-06
Quantity
19 units
Firm Location
Manchester, GA
Official Source
View on FDA website ↗

Reason for Recall

Misbranding; Product may have the wrong device handle which incorrectly identifies the product and also has incorrect deployment depth marker

Distribution

AL, AR, FL, LA, MA, NJ, NY, PA, TN, TX, UT and Italy

Lot / Code Info

Catalog number: 700-103027, Lot number 25952

Root Cause

Other

Action Taken

Consignees were first notified by telephone on 05/02/2006. A follow up letter was sent on 05/03/2006. The European Medical Device Agency, the Italian Competent Authority and British Standards Institution were notified by email on 05/03/2006.