Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog # HG920-1200, HG920-2500 and 12-HG920-144. (Clinical chemistry)

This recall has been terminated (originally issued October 31, 2008).

Company
Pointe Scientific, Inc.
Recall Initiated
September 10, 2008
Posted
October 31, 2008
Terminated
November 27, 2009
Recall Number
Z-0222-2009
Quantity
37
Firm Location
Canton, MI
Official Source
View on FDA website ↗

Reason for Recall

The inability of the product to maintain stated performance specifications through the stated shelf life.

Distribution

Nationwide, Canada, China, Romania and Trinidad.

Lot / Code Info

HG920-1200 lots 727501-021 and 727501-203; HG920-2500 lot 727501-015; and 12-HG920-144 lots 727501-080 and 727501-142..

Root Cause

Process control

Action Taken

Consignees are being notified by letter dated 9/10/08 which will be sent out via fax, email or certified mail to all direct account and drop-ship customers. Customers are advised to dispose of any recalled product found and to consider test results obtained with these lots questionable. Contact Pointe Scientific, Inc. at 1-800-757-5313 or 1-734-487-8300 for assistance.