GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery, Salt Lake City, UT 84116

This recall has been terminated (originally issued January 23, 2008).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
October 18, 2007
Posted
January 23, 2008
Terminated
August 10, 2012
Recall Number
Z-0183-2008
Quantity
169 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field.

Distribution

Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IL, MA, MD, MI, MS, NC, NY, OK, OR, PA, SC, TX, UT, VA, WI, and WV, and countries of Canada, China, France, Hong Kong, Japan, Mexico, Spain, and Switzerland.

Lot / Code Info

Not applicable since this is an arm for all devices listed in product description.

Root Cause

Device Design

Action Taken

Consignees were sent an Urgent Recall Notice via mail dated 10/11/07, notifying the customer to examine the plastic end of the transmitter arm and to not use the transmitter arm, if any cracks or damage are observed. Also to stop using Steam Sterilization method and to utilize the other Sterilization methods indicated in the Operators Manual. The root cause is still under investigation GEHC Surgery will provide, at no charge, a solution to this issue when it becomes available.