Fabius GS Anesthesia Machine. Catalog number 8604699
This recall has been terminated (originally issued October 19, 2006).
- Company
- Draeger Medical, Inc.
- Recall Initiated
- October 3, 2006
- Posted
- October 19, 2006
- Terminated
- August 7, 2007
- Recall Number
- Z-0048-2007
- Quantity
- 429 units
- Firm Location
- Telford, PA
- Official Source
- View on FDA website ↗
Reason for Recall
One of four casters may break loose from chassis
Distribution
Nationwide and Canada.
Lot / Code Info
Catalog number 8604699
Root Cause
Other
Action Taken
The recalling firm issued recall letters to their customers 10/03/2006 to inform them of the problem and that a service represenative will contact them to schedule an inspection of their machines. Should the inspections revealed any affected machines, the machines will be repaired free of charge.