Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.

Company
Diasorin Inc.
Recall Initiated
October 26, 2004
Posted
April 6, 2005
Terminated
July 22, 2006
Recall Number
Z-0674-05
Quantity
7
Firm Location
Stillwater, MN

Reason for Recall

Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.

Distribution

CT, MI, MT, UT & CANADA

Lot / Code Info

software version 12

Root Cause

Other

Action Taken

Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.