RecallDepth

ATO Drill Guide Kit (All Through One Drill Guides Set), model number 07.00860.001, product is packaged in trays.

This recall has been terminated (originally issued October 23, 2004).

Company
Zimmer Spine, Inc.
Recall Initiated
August 24, 2004
Posted
October 23, 2004
Terminated
November 3, 2005
Recall Number
Z-0040-05
Quantity
82 sets
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

To date, there has been one complaint reported. During a Trinica surgical case, the surgeon was provided a standard disposable, Trinica Drill Bit (part number 07.00166.001) for use with the Trinica All Through One Drill Guide Instrumentation, instead of the ALL Through One Drill Bit. This enabled the physician to drill a few millimeters further than expected.

Distribution

Product was distributed nationwide throughout the United States. Product was not distributed outside of the United States.

Lot / Code Info

Serial Numbers: ATO-001 through ATO-0082

Root Cause

Other

Action Taken

The product distributed to accounts will be returned to Zimmer Spine. Product will be quarantined pending investigation and determination of appropriate corrective action.

More recalls by Zimmer Spine, Inc.

View all recalls by this company →