ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In Vitro Diagnostic use only. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests, utilizing photometry and potentiometric technology, and Immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detection technology.

This recall has been terminated (originally issued September 23, 2010).

Company
Abbott Laboratories, Inc
Recall Initiated
October 21, 2008
Posted
September 23, 2010
Terminated
April 13, 2012
Recall Number
Z-2482-2010
Quantity
404 units distributed worldwide (including US)
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

During manufacturing, two components of the ARCHITECT i1000SR were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover.

Distribution

Worldwide distribution:USA including states of: CO, DC, FL, GA, IL, LA, MA, MN, NY, OH, OK, TX, UT, and WV and countries including: Australia, Brazil, Bogota, Canada, China WFOTC, Columbia, Curacao (NETH ANT), Germany, Hong Kong, Japan, Mexico, Singapore, and Taiwan.

Lot / Code Info

Serial Numbers: (1st recall) i1SR01001-1224; 1226-1300; 1302-1306; 1308-1313; 1315-1316; 1318-1352; 1355-1363; 1365; 1376; 1381-1382  Extended to include serial numbers manufactured from 04/01/08 to 03/31/09: i1SR01021, 1026, 1032, 1040, 1036, 1043, 1059, 1065, 1069, 1071, 1071, 1074, 1082, 1094, 1106, 1107, 1109, 1111, 1112, 1114, 1117, 1119, 1121, 1123, 1128, 1133, 1135, 1138, 1139, 1158, 1162, 1167, 1171, 1173, 1177, 1183, 1186, 1188, 1189, 1198, 1200, 1201, 1205, 1209, 1211, 1217, 1225, 1230, 1243, 1246, 1249, 1250, 1260, 1262, 1272, 1277, 1279, 1284, 1286-1288, 1290, 1294, 1301, 1304, 1312, 1317, 1321, 1324-1327, 1332-1335, 1338, 1339, 1342-1345, 1347, 1348, 1351-1356, 1363-1368, 1370-1372, 1375, 1377-1387, 1389-1396, 1398-1410, 1412-1414, 1416-1441, 1443-1530, 1532-1536 (some of these #'s were removed whose field status was Demo/Training and instruments that had TSB 117-004 completed prior to Feb 19.

Root Cause

Mixed-up of materials/components

Action Taken

ABBOTT issued Technical Service Bulletins to Field Service Representatives who then inspected every unit .