RecallDepth

Stellant CT Dual Syringe Kits, Catalog Number SDS-CTP-QFT and SDS-CTP-SPK

This recall has been terminated (originally issued July 9, 2010).

Company
Medrad Inc
Recall Initiated
May 24, 2010
Posted
July 9, 2010
Terminated
October 13, 2010
Recall Number
Z-1954-2010
Quantity
1,382,889 kits
Firm Location
Warrendale, PA
Official Source
View on FDA website ↗

Reason for Recall

Kits contain a T connector that may be susceptible to cracking or breaking or leaking. This could cause product leakage or, less likely, patient injury.

Distribution

nationwide and Worldwide distribution includes AU, BE, BG, BM, BR, CA, CH, CI, CL, CN, CO, CS, DE, DK, EC, EG, FI, FR, GB, GT, HN, IN, IT, JM, JP, KW, KY, MU, MX, MY, NL, OM, PA, PH, PK, PR, SE, SG, TW, YE, ZA.

Lot / Code Info

99725, 99726, 99727, 99728, 99729, 99730, 99731, 99733, 99734, 99747, 99748, 99749, 99750, 99753, 99754, 99755, 99757, 99766, 99767, 99768, 99772, 99775, 99776, 99777, 99778, 99813, 100053, 100464, 100465, 100466, 100467, 100471, 100472, 100473, 100474, 100475, 100476, 100477, 100478, 100479, 100488, 100512, 100513, 100514, 100515, 100516, 100517, 100518, 100519, 100520, 100521, 100523, 100524, 100525, 100526, 100528, 100529, 100531, 100532, 100533, 100535, 100537, 100538, 100539, 100541, 100542, 100543, 100546, 100547, 100549, 100550, 100551, 100552, 100553, 100555, 100558, 100559, 100560, 100561, 100564, 100567, 100568, 100569, 100570, 100897, 101378, 101379, 101380, 101386, 101387, 101391, 101394 with exp dates between 2/2014 to 4/2014.

Root Cause

Other

Action Taken

The recalling firm issued an Urgent Medical Device Field Correction on 5/24/10 requesting customers to check their inventory for the affected lots and complete the fax back form with the amount of product in stock. The manufacturer plans to make arrangements to provide individually packaged low-pressure connector tubing to use with the product in the event that cracking or breaking and leaking of the T connector is observed. This requires vigilance by the customers over the life of all of the devices. For questions concerning this recall, please contact the company 1-800-633-7231, select option 2, and indicate that you are calling about the Stellant CT Dual Syringe Kit field correction.

More recalls by Medrad Inc

View all recalls by this company →