Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB,

This recall has been terminated (originally issued January 5, 2010).

Company
Maquet Inc.
Recall Initiated
November 17, 2009
Posted
January 5, 2010
Terminated
June 4, 2012
Recall Number
Z-0588-2010
Quantity
20837 (7847 of those in US)
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.

Distribution

Worldwide distribution

Lot / Code Info

Serial numbers 00410-24999. Ventilators in the specific range that are marked with one of the following labels are already equipped with the new type ot pressure transducer. 1. Upgrade EM008/02/1; 2. Upgrade SV300 66 82 745  SV900 66 82 747

Root Cause

Device Design

Action Taken

Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.