Ascent Posterior Occipital Cervical Thoracic System, Single Looped Cable with Two Crimps, One Each. Distributed by Blackstone Medical, Springfield, MA. Intended to promote fusion of the cervical spine and occipitothoracic junction.

This recall has been terminated (originally issued March 16, 2010).

Company
Blackstone Medical, Inc.
Recall Initiated
February 2, 2010
Posted
March 16, 2010
Terminated
April 9, 2010
Recall Number
Z-1144-2010
Quantity
25
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

The Ascent Posterior Occipital Cervical Thoracic System Single Looped Cable with two crimps may be made of stainless steel instead of Titanium.

Distribution

Worldwide Distribution -- United States (Texas and California), Australia, Colombia and Greece.

Lot / Code Info

Part Number: 65-2053, Lot Number: 047247.

Root Cause

Labeling mix-ups

Action Taken

Orthofix Spinal Implants contacted customers by phone beginning December 22, 2009 and issued a Notification of Product Removal dated February 2, 2010. Consignees were instructed to Identify, Remove and Return all affected product to the firm. For further information, contact Orthofix Spinal Implants, Regulatory Affairs Department at 1-973-406-2812.