VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Company
Ventana Medical Systems Inc
Recall Initiated
September 28, 2012
Posted
August 6, 2013
Terminated
February 24, 2016
Recall Number
Z-0834-2013
Quantity
54 instruments
Firm Location
Oro Valley, AZ

Reason for Recall

Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu

Distribution

Worldwide distribution: USA (nationwide) and country of: Germany.

Lot / Code Info

Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market).

Root Cause

Device Design

Action Taken

The firm, Roche sent a "Technical Bulletin" and an letter dated October 3, 2012 to its customers. The letter and bulletin described the product, problem and actions to be taken. The customers were informed that Ventana is aggressively working on solutions to this issue. Until mitigations to the overflow condition are available for implementation; inform all personnel in your organization that work with the BenchMark ULTRA or DISCOVERY ULTRA instrument about the potential for fluid overflow to occur during decontamination cycles; follow the instructions provided in the letter to avoid the documented failure mode and follow the instructions provided in the technical bulletin. Contact your local support center with any questions or concerns or call 520-877-7191.