Stryker MIS Bur, 2.0mm Neuro Diamond, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-120.

This recall has been terminated (originally issued January 11, 2010).

Company
Stryker Instruments Div. Of Stryker Corporation
Recall Initiated
July 6, 2009
Posted
January 11, 2010
Terminated
October 25, 2010
Recall Number
Z-0595-2010
Quantity
5438 of all products
Firm Location
Kalamazoo, MI
Official Source
View on FDA website ↗

Reason for Recall

The bur may fracture, resulting in fragments within the surgical site, injury to the patient or surgical team, or both.

Distribution

Nationwide distribution: Australia, Canada, England, France, Hong Kong, Japan, Netherlands, South Africa, Sweden, Switzerland and Taiwan.

Lot / Code Info

All lots distributed 1/30/06 through 3/4/09. Lots 06004017, 06034017, 06075017, 06086017, 06206017, 07002017, 07039017, 07067017, 07121017, 07219017, 07279017, 07320017, 07346017, 08007017, 08030017, 08044017, 08081017, 08205017, 08275017 and 09043017.

Root Cause

Device Design

Action Taken

Consignees were sent a letter dated 7/10/09 instructing them to locate and return product on hand for replacement. Surgeons who use the product were notified via a separate, untitled letter, dated July 13, 2009. A speed warning will be added to the product labels. Questions are directed to the firm at: 1-800-800-4236 ext. 3808.