Alcohol Standard sold by Pointe Scientific; Catalog # 7-A7504-STD.

This recall has been terminated (originally issued September 30, 2006).

Company
Pointe Scientific, Inc.
Recall Initiated
June 9, 2006
Posted
September 30, 2006
Terminated
November 20, 2007
Recall Number
Z-1549-06
Quantity
44 units
Firm Location
Canton, MI
Official Source
View on FDA website ↗

Reason for Recall

Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability.

Distribution

Worldwide distribution ---- including USA and countries of Canada, Ecuador, Germany, Greece, India, Mexico, Peru, Poland, Portugal, Trinidad, United Arab Emirates and United Kingdom.

Lot / Code Info

Lot 509102; exp. 9/06.

Root Cause

Other

Action Taken

Consignees were notified via recall letter dated 6/9/06 and a second letter dated 6/21/06 extending the recall to additional products/lots, to cease using the products, conduct a recall to the user level, destroy products on hand and to notify the firm via the response form of the amount destroyed for replacement purposes.