Skytron General Purpose Surgical Table, battery model; Model 6600B, Dist. by Skytron, Grand Rapids, MI.

Company
Skytron, Div. The Kmw Group, Inc
Recall Initiated
August 4, 2005
Posted
September 13, 2006
Terminated
May 1, 2008
Recall Number
Z-1497-06
Quantity
160 total
Firm Location
Grand Rapids, MI

Reason for Recall

The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop.

Distribution

Nationwide.

Lot / Code Info

All units. Serial numbers 6600B-2D-002 through 6600B-2L-060.

Root Cause

Other

Action Taken

A service bulletin was issued to distributors on 8/12/05 instructing them to identify customers and to retrofit these tables.