Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of 108 pipette tips per rack. Pipette tips are accessory component for Dynex Technologies proprietary automated DSX system.

Company
Dynex Technologies, Inc.
Recall Initiated
June 11, 2004
Posted
July 20, 2004
Terminated
August 18, 2004
Recall Number
Z-1082-04
Quantity
1749 cartons
Firm Location
Chantilly, VA

Reason for Recall

An accessory sampling tip for an in vitro diagnostic test instrument may be inadequately sealed and may cause false assay patient results.

Distribution

Pipette tips were sold to seven OEMs for distribution to end users in the US and internationally.

Lot / Code Info

Lot 412410 printed on carton

Root Cause

Other

Action Taken

The recalling firm notified consignees by faxed letter on 6/11/04 and follow-up letter by mail 6/24/04. The notice advised of vacuum failure associated with lot 412410 of sample tips and recommended quarantine of the defective tips. Consignees were requested to return response form. The 6/24/04 notification recommends that consignees discard/destroy lot 412410 and asks distributors to notify their customers. The firm is working with supplier of the pipette tips to replace recalled product.