RecallDepth

Microflow Needles, catalog #DP8200, with 0 degree tapering, labeled in part ***Non-sterile, Manufactured for Bausch and Lomb, Inc. Rochester, NY 14609***, packaged 6 needles/box.

This recall has been terminated (originally issued October 15, 2003).

Company
Bausch & Lomb Incorporated
Recall Initiated
August 25, 2003
Posted
October 15, 2003
Terminated
December 15, 2003
Recall Number
Z-1279-03
Quantity
22 boxes
Firm Location
Lynchburg, VA
Official Source
View on FDA website ↗

Reason for Recall

The firm's DP8200 Microflow Needle Packs were labeled with the firm's DP8115 Phaco Needle Packs and visa versa.

Distribution

Product was sold to 6 direct accounts in FL, IA, WI, MA and IL. Product was additionally sold to 4 international distributors.

Lot / Code Info

Julian date coded ''2003163''

Root Cause

Other

Action Taken

Firm notified customers by phone and letter 8/25/2003 and advised of needle mix-up. Customers were requested to quarantine affected product and return product with response form.

More recalls by Bausch & Lomb Incorporated

View all recalls by this company →