Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002

This recall has been terminated (originally issued April 28, 2009).

Company
Wright Medical Technology Inc
Recall Initiated
February 19, 2009
Posted
April 28, 2009
Terminated
February 3, 2010
Recall Number
Z-1208-2009
Quantity
682 devices
Firm Location
Arlington, TN
Official Source
View on FDA website ↗

Reason for Recall

The package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated.

Distribution

Worldwide distribution: USA, Canada, Germany, France, Belgium, the Netherlands, Italy, Finland, Austria, Denmark, Norway, Turkey, South Africa, United Kingdom, and Australia.

Lot / Code Info

Lot numbers: 019598209, 038504328, 058486628, 058507733, 058516752, 067422530, 088538652, 098538652, 107439501, 107442247, 117442247, 0712180, S0701050

Root Cause

Process design

Action Taken

Wright Medical notified their distributors of the recall on 03/11/2009 by letter explaining the problem and requesting return of the product from their inventory and from their consignees. The letter included a "fax back" response page to ensure the receipt of the notice. The hospitals and surgeons were notified of the recall also on 03/11/2009. All letters were delivered by Fed Ex.