Monitor portion of the LifeVest device - specifically the Defibrillator printed circuit assembly (PCA) within the monitor. The LifeVest is indicated for adult patients who are at risk for sudden cardiac arrest and are not candidates for or refuse an implantable defibrillator.

This recall has been terminated (originally issued January 28, 2009).

Company
Zoll Lifecor Corporation
Recall Initiated
October 21, 2008
Posted
January 28, 2009
Terminated
November 4, 2009
Recall Number
Z-0883-2009
Quantity
1685 units
Firm Location
Pittsburgh, PA
Official Source
View on FDA website ↗

Reason for Recall

The charging circuit will not shut down properly.

Distribution

Nationwide. OUS: Germany, United Kingdom, Italy, and France.

Lot / Code Info

model numbers WCD3000 and WCD3100

Root Cause

Component design/selection

Action Taken

Physicians with active patients were notified of the product removal by letter via first-class mail dated 10/21/2008. The letter describes the problem and the potential risk. For additional information, contact ZOLL Lifecor Corporation at 1.800.543.3267.