GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Anesthesia systems. The EZchange Module is an optional accessory for these anesthesia systems.

This recall has been terminated (originally issued October 4, 2007).

Company
Datex Ohmeda, Inc
Recall Initiated
July 24, 2007
Posted
October 4, 2007
Terminated
October 16, 2015
Recall Number
Z-0006-2008
Quantity
389,070
Firm Location
Madison, WI
Official Source
View on FDA website ↗

Reason for Recall

Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.

Distribution

Worldwide, including USA, Australia, Belgium, Brazil, Canada, Chile, China, Denmark, Egypt, Estonia, Finland, France, Germany, India, Ireland, Italy, Jordan, Republic of Korea, Kuwait, Mexico, Netherlands, New Zealand, Norway, Peru, Poland, Qatar, Singapore, Slovak Republic, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and Venezuela.

Lot / Code Info

EZchange Manifold with part number 1407-7021-000,  1407-7027-000, M1053007, M1053008.

Root Cause

Vendor change control

Action Taken

Because this issue only occurs when the Disposable Multi Absorber is used in conjunction with EZchange Module, they have contacted all EZchange Module customers. It is critical to contact these customers specifically, rather than all customers shipped Disposable Multi Absorbers as this may cause unwarranted confusion with customers who do not have the EZchange Module. Consignees were sent an letter dated 7/19/2007 describing the product and problem. The letter lists a number of interim actions to implement with the anesthesia systems.