EPIK KNEE SYSTEM Angled Insert Impactor RM/LL; Part #801-03-052 Rev. A; non-sterile; knee prosthesis instrument. Distributed by Encore Medical, L.P., Austin, TX 78758.

This recall has been terminated (originally issued October 10, 2007).

Recall Initiated
August 9, 2007
Posted
October 10, 2007
Terminated
March 10, 2008
Recall Number
Z-0010-2008
Quantity
5 units.
Firm Location
Austin, TX
Official Source
View on FDA website ↗

Reason for Recall

Angled Insert Impactor may scratch Tibial Insert on the articulating surface, causing increased wear at the damaged site.

Distribution

Products distributed to direct accounts in CA, FL and MD.

Lot / Code Info

All lots.

Root Cause

Device Design

Action Taken

Firm notified consignees regarding recall via letter on 08/09/07. Consignees informed that Encore customer service representatives would be contacting them to exchange recalled units with replacements.