ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System.

Company
Diasorin Inc.
Recall Initiated
March 9, 2004
Posted
October 1, 2005
Terminated
July 22, 2006
Recall Number
Z-1576-05
Quantity
128
Firm Location
Stillwater, MN

Reason for Recall

An error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek ELx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay.

Distribution

AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI & WV

Lot / Code Info

version 12 and older software

Root Cause

Other

Action Taken

Dear Customer letters, dated 03/08/2004 were sent to customers informing them of the specification error and instructing users to validate their assays manually.