Synchron LX Systems Version 4.5 Operating Software

This recall has been terminated (originally issued October 6, 2005).

Company
Beckman Coulter Inc.
Recall Initiated
September 6, 2005
Posted
October 6, 2005
Terminated
January 26, 2012
Recall Number
Z-0009-06
Quantity
868
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory's reference interval to below the limit.

Distribution

Nationwide and Canada

Lot / Code Info

Synchron LX 20 and PRO Version 4.5 Operating Software Part Number A21127 (Upgrade Kit) Software Versions V4.5N4.6 CDB (Chemistry Database). Synchron LXi Version 4.5 Operating Software A21128 (Upgrade Kit) Software Versions V4.5N4.6 CDB(Chemistry Database).

Root Cause

Software design

Action Taken

Customers were notified by letter dated September 6, 2005 with work around instruction provided in the letter.