CONTAK RENEWAL 4 AVT CRT-D (model no. M170 and M175) and CONTAK RENEWAL 4 AVT HE CRT-D (model no. M177 and M179). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, 4100 Hamiline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 AVT devices provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies.

This recall has been terminated (originally issued June 30, 2005).

Company
Guidant Corporation
Recall Initiated
June 24, 2005
Posted
June 30, 2005
Terminated
December 31, 2006
Recall Number
Z-0952-05
Quantity
1423
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial tachyarrhythmias is inhibited, while bradycardia pacing is unaffected.

Distribution

worldwide. Including United States, Virgin Islands and countries such as Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Cyprus, Czech Republic, Denmark, Eygpt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Labanon, Luxembourg, Marshall Islands, Mexico, Netherlands, New Zealand, Northern Mariana Islands, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom and Venezuela.

Lot / Code Info

all serial numers

Root Cause

Other

Action Taken

Firm issued a Dr. Doctor letter dated 06/23/05. Letter informs physicians of the issue, clinical implications and recommendations. UPDATE: An Advisory Update, dated 08/01/05 is being issued to physicians to inform them of a new programmer software application upgrade and the programmer screen alerts. Additionally regulatory approval was given for a new switch component that will be incorporated into all Guidant CRT-D products.