Sentinel Implantable Carioverter-DefibrilIator (ICD) model 2010 is a multi-programmable, two-zone, tiered-therapy device that incorporates anti-tachycardia pacing (ATP), cardioversion, and defibrillation (DEF) tachyarrhythmia therapies. The device also includes a programmable Hot Can electrode and bradycardia pacing support.

Company
Ela Medical Llc
Recall Initiated
January 25, 2005
Posted
May 6, 2005
Terminated
July 22, 2006
Recall Number
Z-0801-05
Quantity
38 in US and 47 OUS
Firm Location
Plymouth, MN

Reason for Recall

A change in the automatic capacitor reform schedule during the later portion of the device's useful life can result in excessive 750V charge times and disable the device's ability to report an End of Life (EOL) indicator due to excessive charge times unless the patient is receiving periodic 750V therapy shocks.

Distribution

AZ, CA, GA, IL, IN, KY, MA, MD, MI, MN, NC, NJ, OH, PA and to UK, Italy, Germany and Argentina.

Lot / Code Info

AK630187, AK630131, AK630076, AK630290, AK630173, AK630115, AK730368, AK630130, AK630167, AK630099, AK630106, AK730377, AK730393, AK730443, AK730363, AK630157, AK730442, AK630026, AK630135, AK630102, AK630226, AK630142., AK730431, AK630185, AK630137, AK630230, AK630147, AK630015, AK630030, AK630197, AK630098, AK730448, AK630090, AK630265, AK630160, AK730455, AK730381, AK630146 OUS codes:AK630029, AK630043, AK630045, AK630046, AK630048, AK630055, AK630066, AK630067, AK630069, AK630072, AK630073, AK630075, AK630078, AK630080, AK630083, AK630101, AK630104, AK630108, AK630111, AK630112, AK630121, AK630129, AK630154, AK630155, AK630172, AK630174, AK630175, AK630178, AK630186, AK630199, AK630207, AK630216, AK630221, AK630223, AK630231, AK630247, AK630280, AK630281, AK630283, AK630284, AK630296, AK630297, AK630298, AK630299, AK730435, AK730439, AK730441.

Root Cause

Other

Action Taken

Angeion notified physicians responsible for managing each remaining implanted ICD, also provided them with programming means to change the automatic cap charging schedule, and to monitor and document the reprogramming process.