Peri-Strips Dry with Veritas Collagen Matrix Staple Line Reinforcement (PSD-V), 4506 ETS-V. PSD-V is an implantable surgical mesh comprised of non-crosslinked bovine pericardium in a dehydrated form. Do not reuse. Consult Instructions for Use. Sterilized. Product treated with sodium hydroxide. Produt derived from USDA-inspected cattle. Made in the U.S.A. Rx Only. Synovis Surgical Innovations A Division of Synovis Life Technologies, Inc., 6575 University Ave W., St. Paul, MN 55114-1024 USA. 651 796-7300, 800.487.9627

This recall has been terminated (originally issued June 14, 2005).

Company
Synovis Surgical Inovation Div. Of
Recall Initiated
May 23, 2005
Posted
June 14, 2005
Terminated
November 17, 2005
Recall Number
Z-0891-05
Quantity
320 boxes each box containing 6 pouches
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Device is contained in 2 pouches. The outer pouch's seal may be compromised; this may render the exterior of the interior pouch non-sterile.

Distribution

AK, AL, AZ, CA, DE, FL, GA, MO, OR, TN, TX, VA and WA

Lot / Code Info

Lot Number: 5707464-166925, 5707598-118352, 5708605-151644, 5708622-152524, 5708624-152526, 5709355-167627, 5709825-173142, 5709826-173143 and 5709827-173144

Root Cause

Other

Action Taken

The sites and/or distributors that received the suspect devices were contacted by telephone and given instruction how to return/replace the devices. A Recall Notice, dated 05/31/05, was also sent to customers indicating the model number(s) and quantities of affected product which the institution received and the amount they were to return.