Uroview 2500 X-ray System
This recall has been terminated (originally issued June 21, 2003).
- Company
- Ge Oec Medical Systems, Inc
- Recall Initiated
- May 12, 2003
- Posted
- June 21, 2003
- Terminated
- July 7, 2011
- Recall Number
- Z-0946-03
- Quantity
- 263 units
- Firm Location
- Salt Lake City, UT
- Official Source
- View on FDA website ↗
Reason for Recall
Uncommanded motion of the urological table is possible should the hand control assembly be exposed to liquids.
Distribution
Nationwide. Foreign distribution to Argentina, Australia, Austria, Canada, Germany, Italy, Japan, Mexico, Russia and Switzerland.
Lot / Code Info
Serial numbers beginning with M2 and M9.
Root Cause
Other
Action Taken
A Product Safety Alert letter was sent to all consignees on 5/12/2003 recommending that all hand controls be removed from the units.