RecallDepth

Bulk 16 FR Valved PTFE Sheath/Dilator, Part Number 10614-011

Company
Medamicus, Inc
Recall Initiated
June 6, 2003
Posted
July 2, 2003
Terminated
July 2, 2004
Recall Number
Z-0968-03
Firm Location
Plymouth, MN

Reason for Recall

Stability testing, after accelerated aging of the device, showed that the handle of device did not meet strength specifications.

Distribution

The products were distributed to 4 consignees in Georgia, Missouri, New York and North Carolina. Most of the product was shipped to a distributor in New York which probably shipped to other states. There was no direct distribution to foreign or U.S Government accounts.

Lot / Code Info

Lot Number W24540

Root Cause

Other

Action Taken

The consignees were sent recall letters, dated June 6, 2003 or June 20, 2003, that requested the return of the products.