Zimmer Spine Ardis Inserter, Rx, non-sterile

This recall has been terminated (originally issued December 22, 2012).

Company
Zimmer Inc.
Recall Initiated
November 12, 2012
Posted
December 22, 2012
Terminated
February 25, 2014
Recall Number
Z-0594-2013
Quantity
315 (172 in USA)
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up to 60 minutes could occur. Sharp edges of the broken implant may contact and damage adjacent structures, requiring surgical intervention. Observed health risks

Distribution

Worldwide distribution: USA (nationwide) including states of: AR, AZ, CA, CO, CT, DC, FL, GA, ID, LA, MD, MI, MN, MO, NC, NV, NY, OH, OK, PA, TN, TX, WI, and WV; and countries of: Australia, El Salvador, France and Singapore.

Lot / Code Info

Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71JE, 75LE   Part, 07.00780.701, lot 76CN; part 07.00780.702, lot 76CP;part 07.00900.701, lot 76TR; 07.00900.702, 76TS; 650-0297-01, 70YT; 07.00780.701, 73EC; 07.00780.702, 73EC; 07.00900.701, 76TQ; 650-0239-01, 66PR; 650-0293-01, 70SG ; 650-0293-01, 70SG-R; 650-0179-01, 60CK; 650-0176-01, 60CL; 650-0365-01, 72LP; 07.00780.701, 76CQ; 07.00780.702, 76CR; 07.00780.701, 74EN; 07.00780.702, 74EP; 650-0351-01, 72JH; 650-0351-02, 72JJ; 650-0263-01, 68HL; 650-0208-01, 62FH; 07.00900.701, 76TN; 07.00780.701, 76QG; 07.00780.701, 76LC; 07.00780.702, 76LD;

Root Cause

Device Design

Action Taken

Zimmer Spine, sent a "MEDICAL DEVICE CORRECTION" memo dated November 12, 2012 to all consignees / customers describing product, problem and actions to be taken including revised surgical technique. Beginning the week of December 17, 2012, Zimmer began the immediate retrieval of the affected units as a Class I recall. December 20, 2012 nation wide press outlining the recall was issued. Customers may report adverse events or quality problems experienced with the use of these products to Zimmer Spine by calling 866-774-6368 between the hours of 8AM and 5PM CST or by email at qualitycompliance@zimmer.com.