RayStation Version 2.0.0.15, Radiation Therapy Treatment Planning System. RaySearch Laboratories AB Sveavagen 25, SE-111 34 Stockholm Sweden. support@raysearchlabs.com Designed for treatment planning and analysis of radiation therapy.

This recall has been terminated (originally issued March 26, 2012).

Company
Raysearch Laboratories Ab Sveavaegen 9 Stockholm Sweden
Recall Initiated
December 6, 2011
Posted
March 26, 2012
Terminated
May 11, 2012
Recall Number
Z-1317-2012
Quantity
21 units (16 domestically and 5 internationally)
Official Source
View on FDA website ↗

Reason for Recall

An issue was found with the dose calculation in the RayStation 2.0, that could result in a potential overestimation of the dose. This could lead to the potential under-dosage of a patient.

Distribution

Worldwide Distribution -- US, including the states of FL, MA, NJ, and NY and the country of the Netherlands.

Lot / Code Info

Software version and build number is 2.0.0.15. Model number RayStation 2.0

Root Cause

Device Design

Action Taken

The firm decided to recall and sent Urgent Field Safety Notice Medical Device Correction letters to consignees on 12/06/2011. The letters described the issue, identified the affected product, actions to be taken by the user, a solution, and contact information. Users are to follow the two part workaround, and educate their staff and all users on the workarounds. The letter states that the issue will be resolved in the next version of RayStation. If further information is needed, users are to contact either Freddie Cardel, Director of Customer Support, RaySearch Americas, at 1-877-778-3849 or freddie.cardel@raysearchlabs.com; or Niclas Borglund, Director of Service, RaySearch Laboratories AB, at 46-8-5450-6130 or niclas.borglund@raysearchlabs.com