RecallDepth

MPI Rapid Response HUT Table, Model #RRHUT 1017. The firm name on the label is Medical Positioning Inc., Kansas City, MO. Designed for multiple uses including a conventional examination table, unexplained syncope, neurocardiogenic syncope, chronic fatigue syndrome, and most noninvasive procedures.

This recall has been terminated (originally issued March 15, 2012).

Company
Medical Positioning Inc
Recall Initiated
March 5, 2012
Posted
March 15, 2012
Terminated
July 11, 2013
Recall Number
Z-1211-2012
Quantity
5 tables
Firm Location
Kansas City, MO
Official Source
View on FDA website ↗

Reason for Recall

The warning label regarding electrical shock was not placed on the device warning to have the power switch in the "OFF" position prior to connecting the unit to electric power.

Distribution

Worldwide Distribution-USA (nationwide) including the states of IN, KS, WI, NY, IL, NC, OH, and MS and the country of Thailand.

Lot / Code Info

Serial numbers 802421, 802780, 802781 802783, and 802784

Root Cause

Error in labeling

Action Taken

Medical Positioning Inc. sent a "RAPID RESPONSE HUT TABLE LABELING CORRECTION NOTIFICATION" letter dated March 5, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included an Owner's Manual insert page which clearly states the warning and how to avoid the electric shock hazard and a warning tag to attach to the product's power cord near the plug end using the provided wire tie. Contact the firm at 816-595-2356 for questions regarding this notice.

More recalls by Medical Positioning Inc

View all recalls by this company →