DYNASTY(R) BIOFOAM(R) SHELL, SIZE: 46mm GROUP B, REF DSBFGB46, Rx ONLY, STERILE R, Wright Medical Technology, Inc., hip prosthesis acetabular component. The instrument is used as a hip acetabular shell.

Company
Wright Medical Technology Inc
Recall Initiated
January 3, 2012
Posted
February 14, 2012
Terminated
March 13, 2013
Recall Number
Z-1016-2012
Quantity
160 units
Firm Location
Arlington, TN

Reason for Recall

The possibility exists that particulate debris from the packaging spacer could be unintentionally implanted into a patient and could potentially lead to complications that might necessitate a revision surgery.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Australia, Brazil, Canada, Denmark, and Italy.

Lot / Code Info

Lot Numbers: 069846528, 119925740, 010970893, 020998825, 079853470, 79859628, 079868615, 089870267, 089875761, 089878278, 099890880, 099907005, 109917381, 109921935, 119925740, 119930199, 129937584, 0201014640, 0311295429, 0401072697, 0501102202, 0611368642, 0711367603, 0711367604, 0801144244, 0801150043, 0901165305, 04189088001, 04193758401, 04198097901, 05983919310, 05984293410, 06984652810

Root Cause

Packaging

Action Taken

Wright Medical sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated December 27, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory for the affected product and to return any affected product to the firm. A Product Recall Verification Form was enclosed for customers to complete and return via fax to 901-867-7401. Contact Customer Service at 1-800-238-7117 for questions regarding this recall.