Alere Cholestech LDX Analyzer, Model Number: 10-004, 14-874, 14-875, in vitro diagnostic.

This recall has been terminated (originally issued February 6, 2012).

Company
Alere San Diego
Recall Initiated
November 16, 2010
Posted
February 6, 2012
Terminated
August 22, 2013
Recall Number
Z-0964-2012
Quantity
22,000 meters
Firm Location
San Diego, CA
Official Source
View on FDA website ↗

Reason for Recall

Alere has identified that a bias on results for certain analytes tested on the Alere Cholestech LDX System may exist due to humidity variability.

Distribution

Worldwide distribution including USA

Lot / Code Info

ROM versions prior to 3.40

Root Cause

Device Design

Action Taken

Alere sent an Updated Urgent Recall Notification Letter in March 2012 to all their customers who are in possession of one or more Cholestech LDX Systems. The updated letter provides the customers with important information about the problem identified and the actions to be taken. Customers are instructed to operate their device in a controlled environment that maintains humidity within 40-60% relative humidity if the device has not been upgraded with ROM pack v3.40. If humidity cannot be monitored, the device is no longer CLIA waived. Customers are instructed to complete the customer verification form. A previous notice from Alere sent in November 2010 instructs users to contact Alere to receive the ROM pack upgrade. Customers may email LDXROM@alere.com to request an electronic form to complete and return via email. Customers can expect their ROM packs approximately 10 to 14 business days after Alere receives their request. If customers have transferred ownership to another party, they can email Alere at LDXROM@alere.com with the updated contact information. If customers are not currently using their Cholestech LDX System, it is recommended that the ROM Pack upgrade be installed prior to restarting use of any Cholestech LDX analyzer. Customers with questions regarding this communication can call Alere at (877) 441-7440, option 1 or email at LDXROM@alere.com.

More recalls by Alere San Diego

View all recalls by this company →