MICRO-TOUCH(R), L (indicating size large), NITRILE, Powder Free, Medical Examination Gloves, REF 6034303, 200 Pieces, Manufactured for: Ansell Healthcare - Made in Malaysia
This recall has been terminated (originally issued August 31, 2011).
- Company
- Ansell Healthcare Products Llc
- Recall Initiated
- August 4, 2011
- Posted
- August 31, 2011
- Terminated
- November 27, 2012
- Recall Number
- Z-3116-2011
- Quantity
- 250 cases of 10
- Firm Location
- Dothan, AL
- Official Source
- View on FDA website ↗
Reason for Recall
Gloves were inadvertently released for distribution prior to completion of testing required by the U.S. Food and Drug Administration.
Distribution
Nationwide distribution: NC, GA, IL, MO, LA, and PA
Lot / Code Info
11040194EP
Root Cause
Release of Material/Component prior to receiving test results
Action Taken
Ansell account managers were notified on August 4, 2011 to contact (by phone, fax, e-mail) their respective accounts and notify them of the recall situation. The Ansell account managers were provided with a copy of the recall notification to provide to their accounts. The recall notification includes a request that if these gloves have been further distributed to other customers, please advise them of the recall situation and have them return any outstanding stocks.