Leksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.
This recall has been terminated (originally issued February 14, 2011).
- Company
- Elekta, Inc.
- Recall Initiated
- October 1, 2005
- Posted
- February 14, 2011
- Terminated
- March 2, 2011
- Recall Number
- Z-1223-2011
- Quantity
- 272 units
- Firm Location
- Norcross, GA
- Official Source
- View on FDA website ↗
Reason for Recall
Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
Distribution
Worldwide Distribution
Lot / Code Info
SG33620363, US68537171, SG10821625, 3713G01281, SG03321319, SG33520213, US19502892, C160, SG0320704, SG44120295, 00000000000002, SG31630678, SG24820212, SG24820194, SG44221038, SG34620630, SG40920412, SG31020878, A4986A, SG11520160, SG44720158, SG41720310, SG43820213, 3416G09539, SG24820139, US19257221, SG33820630, 19476882, SG13320399H, SG40520126, SG50425261, SG41720317, US66320311, US19490144, SG21220130, US68154038, US19501776, SW ONLY, SG32520836, 3416G08464, SG21220113, SG44925638, SG04120289, SG51220179, SG2030679, US6824174, SG22920793, SG4820232, 3628G06513, 000000000000019, 00000000000004, US19310946, 3416G05185, US69131975, SG42520530, SG23320693, US68104248, SG23720396, SG24820231, SG14720637, SG43520759, US68513855, SG44820136, SG20220701, SG44120292, US19476874, US6863578, SG04020719, 3416G07810, SG43720660, SG44925635, US68538435, US19322820, SG33320960, US19319754, SG31620624, US10102906, SG44925637, SG43520477, US69172955, SG42720416, 3711C04231, US68154724, SG0620611, SG20220192, SG33620337, US19311023, SG35020182, US19239028, SG50425266, SG51920134, 000000000000011, SG44925633, SG24021335, 0304011939, A4986A, SG34220901, SG1220108, SG32120124, US69132547, SG43720550, US19476879, 6146A06124, SG41920535, 6412A00007, US68500234, US19322145, SG20220694, US68115508, SG41620718, SG10720353, 000000000000014, SG43520485, SG42820297, 000000000000015, SG41020267, US68104599, IE24488098, SG11421969, SG41720337, US6836799, US19344670, SG11720734, US68064889, SG20220679, 000000000000016, SG20220699, SG11421978, SG34220707, US19378673, SG34420200, SG20220143, SG50425271, US10025765, US90345020, US19345857, and SG13420315.
Root Cause
Other
Action Taken
Product bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations.